A Promising Analytical Method has been Crafted, Validated, and Quantified for Estimating Theophylline Utilizing UV Spectroscopic and RP-HPLC Techniques in Conjunction with Stress Degradation Studies to Pinpoint Deterioration

Madhavi, Chundru Sai and Mondal, Sumanta and Chakraborty, Subhadip and Saha, Deyasini and Mondal, Prasenjit (2023) A Promising Analytical Method has been Crafted, Validated, and Quantified for Estimating Theophylline Utilizing UV Spectroscopic and RP-HPLC Techniques in Conjunction with Stress Degradation Studies to Pinpoint Deterioration. Journal of Pharmaceutical Research International, 35 (4). pp. 44-64. ISSN 2456-9119

[thumbnail of Mondal3542023JPRI97538.pdf] Text
Mondal3542023JPRI97538.pdf - Published Version

Download (1MB)

Abstract

Introduction: The current study aimed to develop and assess UV-spectrophotometric (zero order, first order, second order, area under the curve) and RP-HPLC methods for estimating theophylline in its pharmaceutical dosage form.

Methods: A less toxic solvent composition of acetonitrile and 0.1% orthophosphoric acid (25:75 v/v) is used as a mobile phase and diluent for developing both UV-spectroscopic and RP-HPLC techniques. Shimadzu Prominence LC-20A Modular HPLC system with a C18 column (250 × 4.6 mm; 5 µm) was used to develop the RP-HPLC method.

Results: Method A is a zero-order spectrophotometric method for determining theophylline at 271 nm, and the correlation coefficient in the linearity study was found to be 0.9958, LOD, and LOQ are 0.52 and 1.71 µg/mL. Method B is a first-order spectrophotometric method for determining theophylline at 258 nm, and the correlation coefficient in the linearity study was found to be 0.9983, LOD, and LOQ are 0.46 and 1.51 µg/mL. Method C is a second-order spectrophotometric method for determining theophylline at 218 nm, and the correlation coefficient in the linearity study was found to be 0.9941, LOD, and LOQ are 0.38 and 1.25 µg/mL. Method D is an area under the curve spectrophotometric method for determining theophylline at 260 to 282 nm, and the correlation coefficient in the linearity study was found to be 0.9978, LOD, and LOQ are 0.57 and 1.88 µg/mL. Method E is the RP-HPLC method for the determination of theophylline at the retention time of 2.814 min, and the correlation coefficient in the linearity study was found to be 0.9923, LOD, and LOQ are 0.78 and 2.57 µg/mL. Studies on stress degradation show that oxidation and acid degradation mostly impact theophylline solutions.

Conclusion: Theophylline can be determined using the proposed approach, which is convenient, precise, affordable, and reproducible.

Item Type: Article
Subjects: Grantha Library > Medical Science
Depositing User: Unnamed user with email support@granthalibrary.com
Date Deposited: 23 Mar 2023 05:40
Last Modified: 30 Jul 2024 06:38
URI: http://asian.universityeprint.com/id/eprint/420

Actions (login required)

View Item
View Item